Biomechanics Research Today is a free monthly online journal that collates and summarizes the latest research about Biomechanics, including details on mechanics of living organisms, movement, locomotion, prosthetic limbs. | ||||||||
|
Design and development of mechanical embolic protection devices.Eskandari MK Northwestern Memorial Hospital, Division of Vascular Surgery, 10-105 Galter Pavilion, 201 East Huron Street, Chicago, IL 60613, USA. meskanda@nmh.org Carotid angioplasty and stenting (CAS) is rapidly becoming an acceptable alternative to carotid endarterectomy in many patients. Much of the success of CAS is credited to advances in equipment dedicated to this minimally invasive approach. Lower profile delivery systems, tapered stent designs and most importantly, development of mechanical cerebral embolic protection devices (EPDs) have contributed to a reduction in periprocedural neurological complications that now rival results of carotid endarterectomy. Despite a lack of level one evidence in support of EPDs, a meta-analysis as well as two recent institutional reports of CAS with and without EPDs suggest a reduced stroke risk when a mechanical protection device is incorporated as part of the procedure. Since the original description by Theron and colleagues, embolic protection systems are markedly improved in ease of use and effectiveness. The three primary groups of EPDs are distal balloon occlusion, distal filtration and proximal occlusion. Although the ideal EPD has yet to materialize, this review provides insight into current design systems and the accompanying strengths and weaknesses of each. Published 9 May 2006 in Expert Rev Med Devices, 3(3): 387-93.
© 2005-2008 Biomechanics Research Today. All Rights Reserved. |
| ||||||